QCYK Reference material - 700 to 1200 keVHome Page | Examples | 0 to 200 keV | 200 to 700 keV | 1200 to 2000 keV | Calibrations
|
| This spectrum was measured using a 45% n-type detector, mounted in a standard 10cm thick cylindrical graded shield. The source was the QCYK reference material evaporated onto a 5 cm diameter filter paper placed, within a polystyrene counting container, at the centre of the detector face. Such a counting position invites true coincidence summing! In this particular spectrum, there appears to be summing with the germanium escaped X-rays.
|
| Measured Energy (keV) | True energy (keV) | Relative Count rate | Nuclide | DK Mode | Attribution | Em. Prob. (%) | |
| 1 | 813.96 | 814.052 | 0.311 | 88Y | EC | DEP | |
| 2 | 834.85 | 834.838 | 16.946 | 54Mn | EC | FEP | 99.9746 |
| 3 | 898.04 | 898.036 | 21.802 | 88Y | EC | FEP | 93.9 |
| 4 | 902.4 | 902.245 | 0.208 | 88Y | EC | Ka2 sum Ge esc | |
| 902.3122 | 88Y | EC | Ka1 sum Ge esc | ||||
| 5 | 912.44 | 912.134 | 1.417 | 88Y | EC | Ka2 sum | |
| 912.2012 | 88Y | EC | Ka1 sum | ||||
| 6 | 1115.55 | 1115.539 | 12.985 | 65Zn | EC | FEP | 50.6 |
| 7 | 1123.76 | 1123.579 | 0.104 | 65Zn | EC | Ka sum | |
| 1124.479 | 65Zn | EC | Kb sum | ||||
| 8 | 1136.38 | 1137.5296 | 0.054 | 65Zn | EC | Random sum Ag-Ka? | |
| 9 | 1173.25 | 1173.228 | 13.833 | 60Co | b- | FEP | 99.85 |
| Home Page | Appendices | Data | Example Spectra | Links | Spreadsheets | Test Spectra | TOP |